11 Aug 2009 , The Multi-Spectral Source (patent pending) from Santa Barbara Infrared is an integrating sphere based output source with integral extended area blackbody and pulsed laser sources (1.06 µm, 1.57 µm) The MSS system is designed to support testing of gated cameras that are integrated into a multisensor package. The MSS combines both an infrared and laser source, allowing the gated camera to be boresighted with an IR sensor. MSS serves as a uniform backlight for a variety of resolution targets.
- Integrated blackbody source providing IR output controllable in both absolute mode (10°C to 60°C) and differential mode (-15°C to +35C° ΔT)
- Synchronized pulsed laser for precision range testing of gated cameras.
- Simulated range adjustable from 100 to 100,000 meters
- Pulse width adjustable from 20 to 500 ns
- Output pulse radiant energy adjustable from 0.4 to 85 nJ/cm2/sr
|